Manual vacuum aspiration (MVA)

Outpatient uterine evacuation under local anaesthetic, up to ~12 weeks.

Source: NICE NG126 · Ipas

The kit

Aspirator60 ml syringe — creates the vacuum.
Valve buttonsClosed = vacuum held; pressed = vacuum released into cannula.
Plunger armsLock the plunger back to hold vacuum.
Collar stopKeeps plunger in the cylinder.
CannulaeFlexible Karman 4–12 mm; size in mm ≈ weeks of pregnancy.

Used for

Incomplete miscarriageMissed miscarriageRetained productsMolar pregnancyEndometrial biopsy

Steps

  1. Analgesia beforehand (NSAID ± misoprostol priming).
  2. Speculum; clean cervix; tenaculum on anterior lip.
  3. Paracervical block at 4 and 8 o’clock with 1% lidocaine.
  4. Dilate with plastic dilators.
  5. Charge aspirator: close valve, pull plunger until arms lock.
  6. Insert cannula just past internal os; attach aspirator.
  7. Press valve buttons; rotate 180° each way with gentle strokes.
  8. Empty and re-charge if the syringe fills.
  9. Stop when gritty, pink foam, uterus grips cannula.
  10. Float tissue in water to see villi; anti-D only if ≥12+0 weeks and RhD-negative; contraception.
A summary for trained clinicians. Always check the full guideline, the patient in front of you and your hospital's protocol. Spotted an error? Tell us.